Waferpedia

Gaertner

MI108

Metrology & InspectionGaertner MI108 family
Research Quality: 50% complete
GaertnerMI108
No photo available yet

What it is

The Gaertner MI108 is a light-sectioning microscope used for metrology and inspection of surfaces.[1]

The MI108 is an instrument for technical testing of surfaces, and it can be used as a portable unit or incorporated into other holding devices.[1]

How it works

The MI108 uses the Schmaltz concept.[1][2]

The MI108 directs a narrow band of light through an adjustable slit and focuses the beam onto the test piece at a forty-five degree angle.[1]

The MI108 divides the light by the top and bottom surfaces of surface irregularities so the surface geometry can be viewed through the microscope eyepiece.[1]

The MI108 uses a filar micrometer with movable crosshairs to measure the difference between the two beams.[1]

Where it fits in the process flow

General reference — not yet source-verified

The MI108 belongs in surface metrology and inspection rather than in wafer fabrication or device processing.

The MI108 supports inspection after surface generation when the goal is to evaluate geometry, texture, and local features on a part.

Applications

The MI108 is used for studying fit and friction, wear and tear, surface contour, nicks, heights of burrs, depths of grooves, edge profile, thickness of anodized layers, coating thickness, and film profile.[1][2]

The MI108 is used for examining dimensional tolerances and surface quality.[1]

  • Technical testing of surfaces
  • Measurement of surface geometry
  • Study of fit and friction
  • Study of wear and tear
  • Measurement of surface contour
  • Measurement of nicks, burr heights, groove depths, and edge profile
  • Measurement of thickness of anodized layers, coating thickness, and film profile

What do the numbers mean?

Optics & imaging4

Objective magnification
3.48X[1]
Accurate?
Eyepiece magnification
10X[1]
Accurate?
Total magnification
34.8X[1]
Accurate?
Optical working distance
41.2 mm[1]
Accurate?

Configuration & options6

Instrument type
Light-sectioning microscope[1]
Accurate?
Model
MI108[1]
Accurate?
Concept
Schmaltz concept[1]
Accurate?
Light beam angle
45°[1]
Accurate?
Micrometer drum graduation
0.00002 inch per division (20 millionths)[1]
Accurate?
Standard equipment
Base, filtered light source, adjustable slit, filar micrometer eyepiece with external revolution counter, x-y positioning stage with .001 reading micrometer thimbles, fixed angular mount, vertical positioning stage[1]
Accurate?
Interested in this tool?
Need help with this tool?Embed spec card

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What type of microscope is the MI108?

The MI108 is a light-sectioning microscope.[1]

What measurement principle does the MI108 use?

The MI108 uses the Schmaltz concept and measures surface geometry with a filar micrometer.[1]

What light geometry does the MI108 use?

The MI108 uses a beam of light that intersects the surface at a forty-five degree angle.[1]

What are the standard equipment items for the MI108?

Standard equipment includes a base, filtered light source, adjustable slit, filar micrometer eyepiece with external revolution counter, x-y positioning stage with .001 reading micrometer thimbles, fixed angular mount, and vertical positioning stage.[1]

Not publicly documented

The following facts about the MI108 are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the MI108 are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • No publicly documented variants, configuration options, or revision breakpoints of the MI108 are on record.

    Answerable by: an OEM product catalog or an engineer who ordered or specified the tool

  • The control-system platform and OS era of the MI108 are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the MI108 are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the MI108 is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (2)Every fact above is drawn from these public sources
  1. [1]web.archive.org — web.archive.orgweb.archive.org
  2. [2]MI108 — Gaertner | Specs, Sourcing & Service | Waferpedia — waferpedia.comwaferpedia.com
Was this page accurate?

Last updated Oct 5, 2026.

Compare with similar tools