1050 UV-VISversusLambda 750 Spectrophotometer

Perkin Elmer 1050 UV-VIS, Perkin Elmer Lambda 750 Spectrophotometer, side by side. Every value shown is drawn from the cited encyclopedia entries.

Side-by-side comparison of selected equipment models
Attribute
1050 UV-VISPerkin Elmer
OEMPerkin ElmerPerkin Elmer
CategoryPrecision OpticsPrecision Optics
Wafer size——
Process node——
Introduced——
Production run——
Lifecycle——
Lifecycle milestones——
Control system——
Generation——
FamilyPerkin Elmer 1050 UV-VISPerkin Elmer Lambda 750
Cited variants—
Specifications
Asset ID82948[2]—
Last Confirm DateAugust 23, 2024[2]—
ManufacturerPerkin Elmer[2]—
Model1050[2]—
Instrument typeSpectrometer[2]—
UseAnalyzing ultraviolet and visible light absorption spectra[2]—
Operating System—Win7 Professional N[3]
Configuration—UV-VIS + NIR[3]
Principle—Double beam, double monochromator, ratio recording UV/Vis/NIR spectrophotometer with microcomputer electronics, controlled by a DELL PC or compatible personal computer.[4]
Optical System—All reflecting optical system (SiO2 coated) with holographic grating monochromator with 1440 Lines/mm UV/Vis blazed at 240 nm and 360 Lines/mm NIR blazed at 1100 nm, Littrow mounting, sample thickness compensated detector optics.[4]
Beam Splitting System—Chopper (46+ Hz, Cycle: Dark/Sample/Dark/Reference, Chopper Segment Signal Correction CSSC).[4]
Detector—Photomultiplier R955 for high energy in the whole UV/Vis wavelength range; Peltier cooled PbS detector for NIR.[4]
Source—Pre-aligned tungsten-halogen and deuterium.[4]
Wavelength Range—190 nm - 3300 nm[4]
UV/Vis Resolution—≤ 0.17 nm[4]
NIR Resolution—≤ 0.20 nm[4]
Stray Light (at 200 nm, 12 g/L KCl USP/DAP method)—> 2 A ≤ 0.0001 %T[4]
Stray Light (at 220 nm, 10 g/L NaI ASTM method)—≤ 0.0001 %T[4]
Wavelength Accuracy—± 0.15 nm UV/Vis; ± 0.5 nm NIR[4]
Wavelength Reproducibility (UV/Vis, Deuterium Lamp Lines)—≤ 0.1 nm[4]
Wavelength Reproducibility (NIR, Deuterium Lamp Lines)—≤ 0.06 nm[4]
Photometric Accuracy (Double Aperture Method 1 A)—± 0.0012 A[4]
Photometric Linearity (at 3.0 A, addition of filters UV/Vis at 546.1 nm, 2 nm Slit, 1-sec integration)—± 0.020 A[4]
Photometric Range—6 A (using reference-beam attenuation)[4]
Bandpass—0.17 nm - 5.00 nm in 0.01 nm increments UV/Vis range; 0.20 nm - 20.00 nm in 0.04 nm increments NIR range[4]
Photometric Stability (after warm-up at 500 nm, 0 A, 2 nm Slit, 2-sec integration, Peak to Peak)—≤ 0.0002 A/h[4]
Baseline Flatness (190 nm - 3100 nm, slit UV/Vis 2 nm, slit NIR Servo Mode)—± 0.0015 A[4]
Photometric Noise RMS (0 A and 500 nm)—≤ 0.00005 A[4]
Primary Sample Compartment (W x D x H)—200 mm x 300 mm x 220 mm[4]
Secondary Sample Compartment (W x D x H)—480 mm x 300 mm x 220 mm[4]
Instrument Dimensions (W x D x H)—1020 mm x 740 mm x 300 mm[4]
Instrument Weight—~ 77 kg[4]
Digital I/O—RS 232 C[4]
Light Beam (above base plate)—90 mm; 120 mm beam separation; 3 mm - 12 mm beam height[4]
Power Requirements—90 VAC - 250 VAC, 50/60 Hz; 400 VA[4]
Temperature Range—10 °C - 35 °C[4]
Recommended Humidity—10 - 70% relative humidity, non-condensing[4]